Click here for assistance in selecting a Pearson Current Monitor.
Click here for a complete list of specifications for all Standard Pearson Current Monitors.
Click here for information, drawings and specifications on Pearson's Clamp On Current Monitors.
Click here for information on Pearson's custom, built-to-order current monitors.
Click here for a detailed overview of Pearson Current Monitors.
Click here for a detailed review of biasing and Pearson's Current Monitors.
Click here for proceedures to calibrate and test Pearson Current Monitors.
Click here for information and specifications on Pearson Capacitive Voltage Dividers.
Click here for High Voltage Pulse Transformer ordering information.
For new products and news from Pearson Electronics, click here.
Click here for telephone numbers, etc. and to learn more about Pearson Electronics.
To return to the Home Page, click here.
Connect the monitor to a voltage-measuring instrument, using a coaxial cable.

Suitable instruments include oscilloscopes, volt meters, and digitizers. Standard model current monitors have either BNC, UHF or type N output connectors. 50 ohm cables that mate with these connectors are readily available, as are adapters between connector types.

Normally, the monitor is connected to a high-impedance (>1 megohm) input. If the electrical length of the connecting cable is greater than about 1/10 of the wave length for the fastest signal to be viewed, some improvement in pulse-edge fidelity can be obtained by placing a matched load at the instrument end of the cable. In this case, the sensitivity is determined by the parallel combination of the monitor source impedance (usually 50 ohm) and the load impedance. For a 50 ohm source and load, and negligible cable resistance, the sensitivity is divided by two.

Pass the conductor to be monitored through the hole in the monitor, and close the circuit.

The conductor must be insulated adequately for its voltage level. The conductive case of the monitor is painted but not otherwise insulated. The case of single-shielded models, or the inner shield of double-shielded models, is tied to the outer shell of the connector, which is usually grounded to the instrument by the outside conductor of the coaxial cable. The case of double-shielded models is isolated, and can be grounded to the circuit under test through the mounting holes.

Read the measuring instrument.

Convert the voltage reading into a current using the monitor's sensitivity (V/A). Include in the conversion any factor resulting from an external terminating resistance. Remember that the monitor output contains no dc component. For example, a periodic positive pulse current will have a dc component. Since this component is not present in the output signal, the baseline of the pulses will appear to be negative.

Current Monitors
Instruction Guide | Choosing a Monitor
Standard Current Monitors | Clamp On Current Monitors
Custom Current Monitors | Application Notes | Bias Notes
Calibration & Test

Capacitive Voltage Dividerss | High Voltage Pulse Transformers
New Products | Home | Contacting Pearson Electronics

PEARSON ELECTRONICS, INC.
4009 Transport Street, Palo Alto, California, USA 94303
Telephone (650) 494-6444 • Fax (650) 494-6716