Pearson Electronics Current Monitor Instruction Guide Connect the monitor to a voltage-measuring instrument,
using a coaxial cable. Suitable instruments include oscilloscopes, volt
meters, and digitizers. Standard model current monitors have either BNC,
UHF or type N output connectors. 50 ohm cables that mate with these connectors
are readily available, as are adapters between connector types.
Pass the conductor to be monitored through the hole in the monitor, and close the circuit. The conductor must be insulated adequately for its voltage level. The conductive case of the monitor is painted but not otherwise insulated. The case of single-shielded models, or the inner shield of double-shielded models, is tied to the outer shell of the connector, which is usually grounded to the instrument by the outside conductor of the coaxial cable. The case of double-shielded models is isolated, and can be grounded to the circuit under test through the mounting holes. Read the measuring instrument. Convert the voltage reading into a current using the monitor's sensitivity (V/A). Include in the conversion any factor resulting from an external terminating resistance. Remember that the monitor output contains no dc component. For example, a periodic positive pulse current will have a dc component. Since this component is not present in the output signal, the baseline of the pulses will appear to be negative. |
|||