The useable rise time of a current transformer is defined with respect to a transition from one current level to another. If the difference between the levels is defined as 100%, then the  rise time is the interval from 10% to 90% of the transition.

When a fast-rising current transition is applied to a current transformer, it is typical for the output to either overshoot and ring, or rise more slowly than the input current pulse. If the overshoot or ringing amplitude is less than 10% of the transition, the output is considered useable. Therefore, the fastest 10 to 90% transition that causes no more than 10% overshoot and ringing is known as the useable rise time. If the output does not overshoot or ring, then the output voltage rise time becomes the useable rise time.

The figure below illustrates this for a Pearson Model 110. A faster Model 2877 is used to show the input current wave shape on channel 2. The oscilloscope has measured the rise time of channel 2 as 12.90 ns. This is the useable rise time because it has caused a 10% overshoot in the Model 110 output voltage signal shown in channel 1.

In choosing a current monitor, the specified useable rise time should be less than the rise-time of the current pulse to be viewed. 

 

Download Adobe PDF
for engineering documention

Supporting Documents

Particle Accelerator
The Particle Accelerator community uses Pearson Current Monitors for a variety of applications in the area of high . . .
 
Medical Applications
The medical equipment industry uses Pearson Current Transformers in several applications. These include medical. . .
 
Power Industry
The Power Industry uses Pearson Current Transformers to measure transients in switching gear and the observation of. . .
 
EMC Industry
Pearson Clamp-on Current Probes are used to make accurate AC current measurements to meet many of the EMI standards. . .
 
Plasma Research
Pearson Current Probes are used by the Semiconductor and Semiconductor Equipment Industry to make measurements at. . .
 
Capacitive Discharge
Pearson Current Monitors are used to view faithful waveforms and make accurate current measurements from a capacitive. . .
 
Surge Current Testing
Wide Band Current Probes measure surge and in-rush currents for EMC and other industries. A typical requirement. . .
 
Lightning Discharge
Broad Band Current Transformers | Monitors are used to characterize lightning strikes at the Kennedy Space Center and leading. . .
 
Beam Instrumentation
Custom Wide Band Current Monitors are use beam tube applications with our large aperture clamp-on Current Monitors. . .