A review of certain Pearson Current Monitor specifications may assist the engineer in choosing the right model for a particular application.
 

Sensitivity

The considerations here involve the peak current to be measured, the oscilloscope sensitivity, and trade-offs imposed by other specifications.
 

For pulse applications the time domain parameters are:
 

  • Maximum Peak Current This value is based primarily on the voltage-breakdown rating of the connector used. For instance, a 500-volt rating on the connector gives a 5000-ampere peak current rating for a 0.1 volt-per-ampere current monitor.
  • Droop The value listed is the maximum amount to be expected at current levels above a few amperes. At low current levels, low initial core permeability may cause higher droop values and a corresponding increase in the low-frequency -3 dB point for some models.
  • Usable Rise Time If the 10 to 90 percent rise time is greater than the specified usable rise time, initial overshoot and ringing will be less than 10% of the pulse step amplitude.
  • I/t Max This parameter is analogous to the voltage-time constant of a pulse transformer. The product of current times time for a rectangular pulse must not exceed the value listed or the core will saturate, causing a distorted waveform. If two turns are used through the window to obtain twice as much sensitivity, the I/t rating necessarily will be halved.


For the continuous signal applications, the frequency domain parameters are:
 

  • Maximum RMS Current This value is based on heating considerations involving the long-term stability of the internal resistance element in the current monitor.
  • Approximate Low And High Frequency 3 dB Points Due to the ac nature of transformers, the flat midband response will roll off at low frequency. The “corner” or “cut-off” frequency, at which the response is 3 dB down, is specified. Internal resonances determine the useful high frequency cut-off point. Response is within ± 3 dB at the specified high frequency limit.
  • I/f Max This parameter is to sine-wave currents what the I•t product is to rectangular-wave currents. The quotient of peak current divided by frequency must not exceed the listed value or the core will saturate. If two turns are used through the window to obtain twice as much sensitivity, the I/f rating necessarily will be halved.

Supporting Documents

Particle Accelerator
The Particle Accelerator community uses Pearson Current Monitors for a variety of applications in the area of high . . .
 
Medical Applications
The medical equipment industry uses Pearson Current Transformers in several applications. These include medical. . .
 
Power Industry
The Power Industry uses Pearson Current Transformers to measure transients in switching gear and the observation of. . .
 
EMC Industry
Pearson Clamp-on Current Probes are used to make accurate AC current measurements to meet many of the EMI standards. . .
 
Plasma Research
Pearson Current Probes are used by the Semiconductor and Semiconductor Equipment Industry to make measurements at. . .
 
Capacitive Discharge
Pearson Current Monitors are used to view faithful waveforms and make accurate current measurements from a capacitive. . .
 
Surge Current Testing
Wide Band Current Probes measure surge and in-rush currents for EMC and other industries. A typical requirement. . .
 
Lightning Discharge
Broad Band Current Transformers | Monitors are used to characterize lightning strikes at the Kennedy Space Center and leading. . .
 
Beam Instrumentation
Custom Wide Band Current Monitors are use beam tube applications with our large aperture clamp-on Current Monitors. . .